Teledyne-lecroy QPHY-10GBase-T Manual de usuario

Busca en linea o descarga Manual de usuario para Software Teledyne-lecroy QPHY-10GBase-T. Teledyne LeCroy QPHY-10GBase-T User Manual Manual de usuario

  • Descarga
  • Añadir a mis manuales
  • Imprimir

Indice de contenidos

Pagina 1 - Revision A – April, 2014

QPHY-10GBASE-T (Tx and RL Tests) Operator’s Manual Revision A – April, 2014 Relating to the following release versions: Software Version R

Pagina 2

General Setup The first sub-dialog contains general system settings. These remain in effect for each session, regardless of Standard, until changed

Pagina 3 - Table of Contents

QPHY-10GBASE-T Operator’s Manual QualiPHY Test Process Once general system settings are in place, these are the steps for running test sessions. Set

Pagina 4 - About This Manual

Run Reports The QualiPHY software automates report generation. On the framework dialog, go to General Setup > Report to pre-configure reporting

Pagina 5 - Introduction

QPHY-10GBASE-T Operator’s Manual Customizing QualiPHY The pre-loaded configurations cannot be modified. However, you can create your own test configu

Pagina 6

Edit Variables The Variable Setup tab contains a list of test variables. To modify a variable: 1. Select the variable on the Variable Setup tab, t

Pagina 7 - Installation and Setup

QPHY-10GBASE-T Operator’s Manual Edit Test Limits The Limits tab shows the Limit Set currently associated with the configuration. Any limit set can b

Pagina 8 - Set Up Remote Control

X-Replay Mode The X-Replay mode window is an advanced (“developer”) view of QualiPHY. The tree in the upper-left frame enables you to navigate to p

Pagina 9 - Using QualiPHY

QPHY-10GBASE-T Operator’s Manual QPHY-10GBASE-T Tx Testing This section covers the transmitter compliance tests. Test Preparation Before beginning

Pagina 10 - General Setup

QPHY-10GBASE-T Tx Test Descriptions These are the standard 10GBASE-T transmitter compliance tests. Test 1 - Maximum Output Droop The purpose of thi

Pagina 11 - QualiPHY Test Process

QPHY-10GBASE-T Operator’s Manual Figure 6 – Maximum Output Droop Test Results Shown on this screen: • F1 = C1-C2 • P1 = Lvl10ns, = Level@X where X

Pagina 12 - 923663 Rev A

700 Chestnut Ridge Road Chestnut Ridge, NY, 10977-6499 Tel: (845) 425-2000, Fax: (845) 578 5985 teledynelecroy.com © 2014 Teledyne LeCroy, Inc.

Pagina 13 - Customizing QualiPHY

Test 2 - Transmitter Linearity The purpose of this test is to verify that the SFDR is within the conformance limits as defined in section 55.5.3.2

Pagina 14

QPHY-10GBASE-T Operator’s Manual Test 3 - Transmitter Timing Jitter (Master) The purpose of this test is to verify that the RMS period jitter of the

Pagina 15 - Edit Test Limits

Test 4 - Transmitter Clock Frequency The purpose of this test is to verify that the symbol transmission rate of the MASTER PHY transmitter is withi

Pagina 16 - X-Replay Mode

QPHY-10GBASE-T Operator’s Manual Test 5 - Transmitter Power Spectral Density The purpose of this test is to verify that the power spectral density of

Pagina 17 - QPHY-10GBASE-T Tx Testing

Test 6 - Transmitter Power Level The purpose of this test is to verify that the power level of the transmitter is within the conformance limits as

Pagina 18

QPHY-10GBASE-T Operator’s Manual Test 7 - Transmitter Timing Jitter (SLAVE) The purpose of this test is to verify that the RMS period jitter of the S

Pagina 19 - 923663 Rev A 15

QPHY-10GBASE-T Tx Variables There is one variable that can be set for QPHY-10GBASE-T Tx testing: PAIRS TO TEST Specify whether to test a single pai

Pagina 20

QPHY-10GBASE-T Operator’s Manual QPHY-10GBASE-T Return Loss Testing Test Preparation Before beginning any test or data acquisition, the SPARQ network

Pagina 21 - EFAULT PASS CONDITION:

QPHY-10GBASE-T RL Test Configurations All Tests There is only one configuration to choose from. This configuration enables users to perform the fix

Pagina 22

QPHY-10GBASE-T Operator’s Manual DEFAULT PASS CONDITION: The portion of the measured impedance profile that is specified in the standard (0 – 500MHz)

Pagina 23

QPHY-10GBASE-T Operator’s Manual Table of Contents Introduction ...

Pagina 24

QPHY-10GBASE-T RL Variables FIXTURE DE-EMBEDDING CALIBRATION FILE A Specify the full path that points to the L12T file used for de-embedding the fi

Pagina 25

QPHY-10GBASE-T Operator’s Manual Manual Deskewing Procedures This section is only applicable to the oscilloscope and the cables connecting to the osc

Pagina 26 - QPHY-10GBASE-T Limit Sets

• Trigger i. Configure to Source to be FastEdge. ii. Set the Slope to Positive. • Parameter Measurements: i. Set the source for P1 to CX and

Pagina 27 - Test Preparation

QPHY-10GBASE-T Operator’s Manual 10. Save Channel X to M1. • Click File  Save Waveform. • Set Save To Memory. • Set the Source to CX. • Set the

Pagina 28 - QPHY-10GBASE-T RL Tests

Cable Deskewing Without Using the Fast Edge Output The following procedure demonstrates how to manually deskew two oscilloscope channels and cables

Pagina 29

QPHY-10GBASE-T Operator’s Manual In the previous procedure, we used the default setup of the Skew parameter (which is detecting positive edges on bot

Pagina 31 - Manual Deskewing Procedures

Table of Figures Figure 1 - QualiPHY framework dialog and Standard selection menu. ... 5 Figure 2 -

Pagina 32

QPHY-10GBASE-T Operator’s Manual Introduction About QualiPHY QualiPHY is a highly automated compliance test software meant to help you develop and va

Pagina 33 - 923663 Rev A 29

Required Host Computer System Usually, the oscilloscope is the host computer for the QualiPHY software, and all models that meet the acquisition r

Pagina 34

QPHY-10GBASE-T Operator’s Manual Installation and Setup QualiPHY is a Windows-based application that can be configured with one or more serial data c

Pagina 35 - Error Messages

Set Up Remote Control QualiPHY software can be executed from a remote host computer, controlling the oscilloscope through a LAN Connection. To set

Pagina 36

QPHY-10GBASE-T Operator’s Manual Using QualiPHY This section provides an overview of the QualiPHY user interface and general procedures. For detaile

Comentarios a estos manuales

Sin comentarios