Electrical Telecom Pulse Mask Test Software
Add Measurements to Mask Test
In addition to measuring pulse compliance relative to a given mask, a full set of parametric measurements
are available. Measurements are made on the masked waveform as it appears on screen.
NOTE: Some measurements are incremental and will be reset if you Clear Sweeps during the test.
1. From the menu bar, choose Measure > Measure Setup and choose a parameter location (Px).
2. Touch Source and select ET.
3. Touch Measure and select the desired measurement parameter. If right-hand dialogs offer additional
settings for the selected parameter, enter values on these dialogs.
The measurement read out appears in a table beneath the grid.
4. To place markers exactly where on the waveform measurements are being made, choose a Marker type
and check Always On in the Help section of the dialog.
You can repeat this setup with different parameters for as many locations as your instrument offers.
6
923136 Rev A