QPHY-SAS3 TEST DESCRIPTIONS
Tx Group 1 Out of Band Tests 5.1.x (OOB)
There are 4 tests run in this group. The tests that are run are:
1. 5.1.1 – Tx Maximum Noise During OOB Idle
2. 5.1.2Tx OOB Burst Amplitude
3. 5.1.3Tx OOB Offset Delta
4. 5.1.4Tx OOB Common Mode Delta
After the completion of the Out of Band Tests the oscilloscope is in the following configuration:
Figure 9 - Oscilloscope Configuration after Out of Band Tests
Shown on this screen:
F3 is the Idle Differential Signal. This trace represents only the idle portion of the differential OOB
signal.
F4 is the Burst Differential Signal. This trace represents only the burst portion of the differential OOB
signal after it is filtered by a 4.5GHz low-pass Butterworth filter.
Having these two traces separate from each other allows us to perform the measurement on the idle
portion and the burst portion separately.
F5 is the Burst Common Mode signal.
F6 is the Wide Band AM Demodulated Differential Signal. This is used for the separation of the idle
portion from the burst portion.
F7 is the Idle Common Mode signal. This is calculated by summing the two inputs and dividing them
by 2.
922545 Rev A 19