ZiOscilloscopeSynchronizationModule
2
91824600RevA
TheuserinterfaceforcommontriggeringanddisplayofchannelsisenabledontheMasteroscilloscopewhen
theZi8CHSYNCHhardwareProBusmoduleisattachedtotheMasterAUXOUTconnector.Alltriggeringis
performedbytheZi8CHSYNCHmodule;datatransferisviaanEthernetcrossovercableconnected
tothe
MasterandSlaveEthernetports,orthroughanetworkconnection .Handshakingbetweenoscilloscopesis
automaticoncetheIPaddressoftheSlaveoscilloscopeisdefined.
Connections
Thegeneralsetupinvolvesthefollowing:
TheZi8CHSYNCHmoduleisconnectedtotheAUXOUTProBusconnectorontheMasteroscilloscope.
Themodulecontainsacomparatorwithaprogrammablethresholdandabuffertoacceptanexternal
customertriggersignal(inUserEdgetriggermode)witha
relativelyslowrisetimeandoutputstwo
identicaltriggersignals(TRIGOUT1andTRIGOUT2)with~100psrisetimetotheMasterandSlave
oscilloscopesAUXINProBusconnectors.Theseconnectionsserveastheircorrespondingsynchronized
triggerinputs.Ifnoexternalcustomertriggersignalisavailable,themodulecanbe
selftriggeredina
Periodictriggermode.
Finally,theSLAVEMASTERSYNCHconnectionismadefromtheSlaveAUXOUTtothemodule.
TheSLAVEMASTERSYNCHconnectionprovidespositivefeedbackfromtheSlavetotheZi8CHSYNCH
modulethatanacquisitionhasoccurredontheSlave(ensuringMaster
+Slaveacquisitionintegrity).This
isprovidedviaaTRIGENABLEDoutputsignalfromtheSlaveoscilloscope.The10MHzReferenceClock
OutontheMasteroscilloscopeisconnectedtothe10MHzReferenceClockInontheSlaveoscilloscope
inordertosynchronizethetwooscilloscopeacquisiti ons.Uponproper
connectionandreceiptofavalid
TRIGINsignal,theMasterandSlaveoscilloscopesimultaneouslyacquiredata.