Teledyne-lecroy AORM - Advanced Optical Recording Measurements Manual de usuario Pagina 14

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 125
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 13
12 ISSUED: June 2013 923133 Rev A
Field Description
Show
Determine
s which table of values is displayed below the grid.
When Parameters is selected, a checkbox is displayed that
enables you to turn on or off the display of statistics values.
Source This can be a channel, math function, or memory
trace. A Use Equalizer checkbox is provided to
apply a filter to the data source.
Data Gate This can be a channel, math function, memory
trace, or other. It specifies the input that will be
used to determine where to perform measurements
on the input signal. The polarity is always high.
Use
Equalizer
This checkbox enables filtering of the data source.
Data Slope Polarity of the pits/spaces to use for the
measurement, when appropriate. Pos polarity
refers to pits, Neg polarity refers to spaces, Both
can be selected to use both pits and spaces.
Data
Hysteresis The Hysteresis selection imposes a limit above and
below the Threshold, which precludes
measurements of noise or other perturbations
within this band. The width of the band is specified
in divisions.
Guidelines for Use
1. Hysteresi
s must be larger than the ma
ximum
noise spi
ke you want to ignore.
2. The larg
est value of hysteresis usable
is less
than the dista
nce from the threshold to the closest
extreme value of the waveform.
3. Unle
ss you know the largest noise and closes
t
extreme level
that will ever occur on any cycle,
leave some margin on both sides of the threshold.
Vista de pagina 13
1 2 ... 9 10 11 12 13 14 15 16 17 18 19 ... 124 125

Comentarios a estos manuales

Sin comentarios