AORM Software Package
923133 Rev A ISSUED: June 2013 89
Further zooming clearly
identifies the problem: a
burst error that prevented
the positive polarity width
that starts at 859.61
µ
s
from reaching its peak
value. This defect
caused the reflectivity to
drop and to erratically
fluctuate throughout the
duration of the burst
error. The defect affected
the width of the next pit
as well, which created
the pit width centered
between the 4T and 5T
band of the XY plot.