AORM Software Package
923133 Rev A ISSUED: June 2013 21
4. Touch inside the Operator1 field and select Trend
from the Select Math
Opera
tor pop-up menu. The "Trend" setup dialog will appear at the right of the screen
:
5.
Decide whether all the parameter values, all per trace, or only the averag
e of all
para
meter calculations for each waveform acquisition should be placed in the tre
nd.
All -- every p
arameter calculation on each waveform will be placed
in the trend.
Av
erage -- trends only the average of all the values calculated on a given acquisition an
d
yields on
e point in the trend per acqui
sition.
All per Trace -- for ea
ch acquisition, clears the buffer and places all param
eter
cal
culations from the new data in the trend. Unless this is specifically required, All shou
ld
be sel
ected.
6.
Choose the number of values to be placed in the generated trend.
7.
If desired, you can also configure the center and heig
ht of the trend in the base units of
the para
meter being trended. However, this is not a requirement, and Find Scale can be
used to center the trend after it has been
calculated.
Cen
ter is for selecting the mantissa, exponent, or number of digits resolution, us
ing the
asso
ciated knob. The configuration is the value at the horizontal center line on th
e grid,
while u
nits are those of the parameter trende
d.
Height/di
v selects the vertical value of each vertical screen division. Units ar
e those of
the para
meter trend
ed.